Calibration for Hand-held Browser Probe Heads

Calibration of the hand-held browser probe heads consists of a vertical calibration
and a skew calibration. The vertical calibration should be performed before the
skew calibration. Both calibrations should be performed for best probe
measurement performance.
Before calibrating the probe, verify that the Infiniium oscilloscope has been calibrated recently
N OT E
and that the calibration Δ temperature is within ±5 °C. If this is not the case, calibrate the
oscilloscope before calibrating the probe. This information is found in Infiniium Calibration
dialog box.
Calibration Setup
The calibration procedure requires the following parts.
• BNC (male) to SMA (male) adapter
• Deskew fixture
• 50Ω SMA terminator
1
2
3
4
5
6
7
8
9
10
11
12
1130B-Series Probes User's Guide
As shown in
Figure 54
to the deskew fixture on the connector closest to the yellow pincher.
Connect the 50Ω SMA terminator to the connector farthest from the yellow
pincher.
Connect the BNC side of the deskew fixture to the Aux Out of the Infiniium
oscilloscope.
Connect the probe to an oscilloscope channel.
Place the positive resistor tip of the browser on the center conductor of the
deskew fixture between the green line and front end of the yellow pincher. The
negative resistor tip or ground pin of the browser must be on either of the two
outside conductors (ground) of the deskew fixture.
On the Infiniium oscilloscope in the Setup menu, select the channel connected
to the probe.
In the Channel Setup dialog box select the Probes... button.
In the Probe Setup dialog box select the Calibrate Probe... button.
In the Probe Cal dialog box select the Calibrated Atten/Offset radio button.
Select the Start Atten/Offset Calibration... button and follow the on-screen
instructions for the vertical calibration procedure.
Once the vertical calibration has successfully completed, select the Calibrated
Skew... button.
Select the Start Skew Calibration... button and follow the on-screen
instructions for the skew calibration.