To properly trim the probe head's lead wires
1
Figure 18 MX0100A Trim Gauge Template (MX0100-94302)
2
3
Soldering an MX0100A Probe Head to DUT
To solder the probe tip lead wires to DUT
1
1130B-Series Probes User's Guide
Use the Keysight supplied trim gauge template that is included as part of the
MX0100A probe head kit.
Using tweezers, place the lead wire over the outline of the lead wires as shown
on the trim gauge template. The trim gauge template displays two lengths:
3.4 mm and 1.5 mm. Choose the correct length as per your DUT.
Using the cutting tweezers, trim the lead wires even with the trim lines.
You can spread the probe head's lead wires within the range of 0mm to 7mm
N OT E
span without causing any significant variation in its available bandwidth.
The tools included in the MX0102A soldering toolkit can be of great use
N OT E
while soldering the MX0100A probe head to DUT (see
purchase this toolkit separately.
Trim the length of the MX0100A probe head lead wires to match your DUT's
geometry (see
page
40). You may use the cutting tweezers (Keysight part
number 8710-2838) included in the Soldering toolkit.