Using Probe Heads
To check the MX0100A probe head for any damage
You can use a Digital Multimeter to check the resistance measurement of your
MX0100A probe head. If the resistance measurement between the probe head's tip
and tail is 25.2 kohm, then the probe head is usable,
Figure 22 Resistance measurement for an undamaged MX0100A probe head
For a damaged probe head, the resistance measurement between the MX0100A
probe head tip wire and the center conductor of the SMP connector of the
MX0100A is displayed as Infinite.
Figure 23 Resistance measurement for a damaged MX0100A probe head
Replacing an MX0100A Probe Tip Lead Wire
Use the following procedure to install or replace the lead wires on the MX0100A
probe head in the event of damage or break off due to use. Depending on your
probing application, you can order either 9 mil or 10 mil wire as listed in the
following table. These wires are also included in the Keysight MX0102A Soldering
Toolkit (see
46
page 29
for details).
Table 10 Required Wire Types
Wire Type
Regular Solder Wire (lead free)
Requires standard lead-free soldering
temperatures (330 °C to 350 °C).
(NOTE: This alloy melts at 217
Wire Diameter
.009" diameter
o
C.)
Part Number
MX0102-21302