Keysight Technologies 1130B User Manual page 209

Differential and single-ended probes
Table of Contents

Advertisement

Figure 227 SMA Cable Connected to Port 2
13
14
15
16
1130B-Series Probes User's Guide
Get the opposite sex of the Calibration Standards for the next step.
Perform Calibration for the PORT 2 side of the Reference plane.
Press the reflection screen key.
a
Connect the open end of Calibration Standard to the available end of the
b
PORT 2 SMA cable.
Select the open screen key under the Reverse group.
c
Wait until the VNA beeps indicating that it has completed the task.
d
Connect short end of Calibration Standard to the available end of the PORT 2
e
SMA cable.
Select short screen key the Reverse group.
f
Wait until the VNA beeps indicating that it has completed the task.
g
Connect load end of Calibration Standard to the available end of the PORT 2
h
SMA cable.
Select the loads screen key the Reverse group.
i
Press broadband screen key selection.
j
Wait until the VNA beeps indicating that it has completed the task.
k
Press the done loads screen key.
l
You have just calibrated the other side of the reference plane.
m
Press standards done key.
Connect PORT 2 SMA cable to the non-pincher side of PV Fixture.
Performance Verification
209