Differential and single-ended probes
Table of Contents

Advertisement

3
Calibrating Probes
Figure 50 Good Connection
Figure 51 Bad Connection
Step 3. Running the Probe Calibration and Deskew
1
2
3
4
5
82
On the Infiniium oscilloscope in the Setup menu, select the channel connected
to the probe.
In the Channel Setup dialog box select the Probes... button.
In the Probe Setup dialog box select the Calibrate Probe... button.
In the Probe Cal dialog box select the Calibrated Atten/Offset radio button.
Select the Start Atten/Offset Calibration... button and follow the on-screen
instructions for the vertical calibration procedure.
1130B-Series Probes User's Guide