Keysight Technologies 1130B User Manual page 40

Differential and single-ended probes
Table of Contents

Advertisement

2
Using Probe Heads
Required Infiniium Software Version
The MX0100A probe head requires the Infiniium software version 6.3 or higher.
Table 9
MX0100A Probe Head Kit Components
Component
Micro Probe Heads (with pre-wired probe tips) * 5
Probe Tip Wire (.004" diameter)
(To make ground connections)
Bullet Adapter
Trim Gauge Template (see
Figure
* Quantity varies based on the purchased option.
If the probe tip lead wire is damaged or worn out, replace the lead wire using the procedure
N OT E
"Replacing an MX0100A Probe Tip Lead Wire"
To know how to avoid damage to the MX0100A probe head, refer to the topic
Probe Head Handling Precautions"
Trimming the Lead Wires of MX0100A Probe Head
Before soldering, trim the probe head's lead wires matching your DUT's geometry.
You can choose from the following lead wire lengths:
• 135 mil (3.4 mm) - The probe head is shipped with this factory-trimmed standard
• 60 mil (1.5 mm) - If your DUT's geometry allows you to use shorter lead wire
40
Option 001
1 wire spool
1
18)
1
length. Use this lead wire length to accommodate variable-pitch targets. With
this length, you get the maximum convenience in terms of longer reach and the
available bandwidth is the full bandwidth of the probe amplifier being used
(1134B: 7 GHz, 1132B: 5 GHz, 1131B: 3.5 GHz, and 1130B: 1.5 GHz).
length, trim the wires to this length to get the maximum performance. Use this
lead wire length to accommodate small fine-pitch targets. The available
bandwidth is the full bandwidth of the probe amplifier being used (1134B: 7
GHz, 1132B: 5 GHz, 1131B: 3.5 GHz, and 1130B: 1.5 GHz).
You need to specify your choice of lead wire length (3.4 mm or 1.5 mm) in
N OT E
the Probe Configuration dialog box of the Infiniium software GUI. This allows
the software to load the appropriate s parameter file applicable to that wire
length. The s parameter file adjusts the frequency response to enhance the
measurements accuracy.
Quantity *
Option 002
Option 003
25
50
5 wire spools
10 wire spools
5
10
5
10
on page 46".
on page 43".
Part Number
MX0100A
MX0102-21301
MX0103A
MX0100-94302
"MX0100A
1130B-Series Probes User's Guide

Advertisement

Table of Contents
loading

This manual is also suitable for:

1131b1132b1134b

Table of Contents