Keysight Technologies 1130B User Manual page 212

Differential and single-ended probes
Table of Contents

Advertisement

9
Performance Verification
Measuring Vout Response
38
39
40
41
42
43
Figure 231 Typical V
Displaying the Vout/Vin Response
44
45
212
Disconnect the PORT 2 cable from PV/DS test board and attach to probe output
on the AutoProbe Adapter.
Connect the Calibration Standard load to PV/DS test board (non-pincher side).
Press [scale reference] key on the VNA.
Set reference value to –20 dB.
Hold probe in place as described previously.
The display on screen is V
Figure 231
on page 212.
Waveform for an 1134B Probe
out
Press the [Display] key.
Then select the Data/Memory screen key. You may need to adjust the Reference
Value, located under the Scale Ref key, slightly to position the waveform at
center screen. The waveform should be similar to that shown in
page 213.
and it should be similar to that shown in
out
Figure 232
1130B-Series Probes User's Guide
on

Advertisement