Step 2. Verifying the Connection
1
2
3
4
If you see a waveform similar to that of
and should check all of your probe connections.
Figure 49 Connecting the Probe and Deskew Fixture
1130B-Series Probes User's Guide
On the Infiniium oscilloscope, press the autoscale button on the front panel.
Set the volts per division to 100 mV/div.
Set the horizontal scale to 1.00 ns/div.
Set the horizontal position to approximately 3 ns. You should see a waveform
similar to that in
Figure 50
InfiniiMax Probe
BNC to SMA Connector
Pincher
Deskew Fixture
50W SMA Terminator
on page 82.
Figure
51, then you have a bad connection
3
Calibrating Probes
81