Performance Verification
Figure 226 PV Fixture Connected to VNA
11
12
208
Perform a calibration at the reference plane:
Press the reflection screen key.
a
Connect the open end of 85052D Calibration Standard to the non-pincher
b
side of the PV/DS test board.
Select the open screen key under the Forward group.
c
Wait until the VNA beeps indicating that it has completed the task.
d
Connect short end of Calibration Standard to the non-pincher side of the
e
PV/DS test board.
Select short screen key under the Forward group.
f
Wait until the VNA beeps indicating that it has completed the task.
g
Connect load end of Calibration Standard to the non-pincher side of the
h
PV/DS test board.
Select the loads screen key under the Forward group.
i
Press broadband screen key selection.
j
Wait until the VNA beeps indicating that it has completed the task.
k
Press the done loads screen key.
l
You have just calibrated one side of the reference plane.
m
Connect the other high quality SMA cable to the VNA's PORT 2 connector.
1130B-Series Probes User's Guide