11. Sterilization

Only the lower piston assembly of LABNET Excel electronic pipettor can
be steam-autoclaved (121
to exceed the the temperature of 121
autoclaving cycle.
period of time will damage the manifold and void the warranty. The
autoclaved parts must be allowed to dry completely at room temperature for
at least 2 hours before reassembly. Users can follow Figures 11~14 to
disassemble the LABNET Excel electronic pipettor. Please "DO NOT" use
excessive force to pull down the lower part before releasing the cylinder
lock as shown in Figure 10. This action will break the connection
mechanism of the lower part, permenently damaging the instrument and
will void the warranty.
Don't use force to pull the lower portion before
release the cylinder lock.
Figure 10: Incorrect removal of the lower piston assembly.
o
C, 1 bar, 20 minutes). NOTE: It is important not
o
C or 20 minute time during the
Autoclaving at a higher temperature or for longer
26
Cylinder lock