10.1. SA (Sequential Aspirating) Mode

Not available on 1200ul models because the diameter of 1000/1200ul tips
does not allow the required air gap to be maintained.
The LABNET Excel electronic pipettor performs repetitive aspirations of
the selected volumes. An air gap will be created automatically between two
aspirations. When the sequential aspiration is completed, the next operation
will dispense all of the aspiration volume together.
1. Press
«EXIT» to enter function mode selection.
2. Press
«Up/Down» repeatedly until the
3. Press
«SET» to confirm the selection.
4. Use
«Up/Down» until the
5. Press
«SET» to confirm the selection.
6. Under the
set the volume of the first aspiration.
7. Press
«SET» to confirm the selection.
8. Repeat steps 6 and 7 to set the sequential aspiration volumes. The total
aspirating volume (including air gap volume) cannot exceed maximum
volume range.
9. Press
«SET» to confirm the selection.
10. The count "01" and aspirating volume will appear on the display. The
pipetting direction and speed
electronic pipette is ready for liquid pick-up. Press the
button to aspirate the samples.
an air gap (by pressing the
sample) between sample aspirations.
symbol starts flashing.
symbol, "01" will appear. Press
will flash to indicate that the LABNET
will appear to remind users to create
button with the tip out of the
PLUNGER
23
symbol starts flashing.
«Up/Down» to
PLUNGER