11. When the aspirations are completed, the LABNET Excel electronic
pipettor will make two beeping sounds and
the LABNET electronic pipette is ready for dispensing.
12. After the sequential aspirating cycle has been completed and the liquid
has been dispensed, pressing the
whole cycle again until users change the function mode.
Note:
1. If the blow-out setting is selected, the LABNET Excel electronic
pipettor will automatically blow-out the liquid during sequential
aspiration mode.
2. 1200ul models do not include the Sequential Aspirating (SA) function,
because the air gap is not strong enough to support the weight of
samples.
3. Sequential Aspirating (SA) function will not perform Reverse Pipetting
even if it is selected in System (SYS) mode.
10.2. SD (Sequential Dispensing) Mode
The LABNET Excel electronic pipettor performs repetitive dispensing of
various selected volumes. As with Multiple Dispensing mode, for best
accuracy in SD Mode, it is recommended to select Reverse Pipetting in
System Mode.
1. Press
«EXIT» to enter function mode selection.
2. Press
«Up/Down» repeatedly until
3. Press
«SET» to confirm the selection.
4. Press
«Up/Down» repeatedly until
will flash to indicate
button will re-activate the
PLUNGER
symbol starts flashing.
symbol starts flashing.
24
Need help?
Do you have a question about the Excel E10-1 and is the answer not in the manual?
Questions and answers