Step
Action
Click the Start button to begin exposure.
The index view and expanded view can be switched by clicking the Index or Image tab.
Note:
Index view
Exposure automatically stops after the number of images set in Repeat Limit.
Note:
TIP:
7
TIP:
ImageQuant LAS 4000 User Manual 28-9607-42 AB
The displayed expanded image is the latest image.
Clicking the Stop button cancels the current exposure.
It is possible to stop exposures automatically when the image is saturated. This
function is activated in Edit:Preference under the Increment tab.
During exposure, the gradation conversion curve can be changed and the grada-
tion adjusted.
When the Apply All button is clicked, the adjusted gradation will be applied to all
images.
1
2
3
4
No.
Description
1
The range can be changed by
dragging the mouse.
2
The gradation conversion
curve can be toggled between
Linear or Sigmoid.
3
The gradation can be
changed by dragging the
mouse.
4 Operating the ImageQuant LAS 4000
Image view
5
No.
4
5
4.10 Exposing consecutively (Increment)
Description
The changed gradation is ap-
plied to all images.
Return to initial setting.