The LEDs of the battery status display flash successively.
»
One LED is continuously illuminated. This indicates the current battery status.
»
The battery is fully recharged when the top red LED is continuously illuminated.
»
Recharging via the wall and device support
If you have attached the charging accessories to a wall and device support then the
device will be charged automatically: Chapter „8.1 Wall and device support" on page
50.
1. Attach the device to the wall and device support.
The LEDs of the battery status display flash successively.
»
One LED is continuously illuminated. This indicates the current battery status.
»
The battery is fully recharged when the top red LED is continuously illuminated.
»
3.4 Connection and removal of canister system and
3.4.1 DDS canister system
WARNING
Risk of infection by contaminated bacterial filter and canister lid.
Deadly diseases can be transmitted.
•
Do not operate the device without a bacterial filter. We recommend you always to
store at least one spare bacterial filter.
•
Always wear disposable gloves when changing the bacterial filter.
•
Prior to each use please check whether the bacterial filter is dry and clean. Replace
the bacterial filter with a new filter if it is discoloured, contaminated or oversucked.
A bacterial filter may never be reused.
•
Exchange the bacterial filter when changing the patient. ATMOS recommends:
Replace the bacterial filter after 14 days, even if there is no patient change.
18
Setting up and starting up
hoses