These operating instructions contain important notes on how to operate the ATMOS A / C 161 / 261 Aspirator safely, correctly and effectively. They help to avoid risks, and also to reduce repair costs and down-times. This increases, among other things, the reliability and service life of the device.
1.2 Function Contraindications:Not suitable for: • continuous operation for drainage procedures in the low The ATMOS A / C 161 / 261 Aspirator is an exceptionally vacuum range (e.g. cardiothoracic or wound drainage) handy small suction device. • permanent endoscopic use It is driven by an electromotive, maintenance-free swing •...
1.0 Introduction 1.4 Scope of supply • Prior to dispatch, this ATMOS device was subjected to an extensive functional test and has been carefully packed. Nevertheless, please compare the contents of the shipment on completeness immediately upon receipt (see delivery note).
Medical Devices Act. • Prior to starting up the ATMOS A / C 161 / 261 Aspirator, read these operating instructions carefully. • The electrical outlet is in fact carried out three-pole, but the middle contact pin normally ground wire connection) •...
3.3 Starting up • The ATMOS A / C 161 / 261 Aspirator is delivered ready for use. • Remove the device from the packaging. Check whether the voltage on the type plate correspond to the mains power supply.
4.2 Set vacuum • Connect the suction hose () to the suction catheter () • Only for the ATMOS C 161 / 261 Aspirator by means of the fingertip (not included in the scope of Set the desired vacuum by closing the fingertip () and delivery) ().
5.0 Operation DDS Important notes on safety for the viral filter must not be dried and not reused. DDS secretion canister system • Replace the DDS bacterial and viral filter after a patient change. Replace the DDS bacterial and • Never operate the device without a viral filter after 14 days, even if there is no patient DDS bacterial and viral filter.
6.0 Operation Receptal ® 6.1 Support for Receptal secretion canister ® Insert the thread of the support from above into the borehole at the bottom side of the device. Screw it together with the supplied screw nut. Fig. 11. 6.2 Assembling the Receptal secretion canister system ®...
7.0 Operation Medi-Vac and Serres ® ® 7.1 Support for the Medi-Vac ® Medi-Vac secretion canister ® Insert the thread of the support from the top into the borehole at the bottom of the device. Screw it together with the supplied screw nut. 7.2 Assembling the Medi-Vac secretion canister system...
Please note that the parts should be rinsed with clear a certified ATMOS partner or a specialist authorised by water for at least 10 seconds in order to achieve a better ATMOS in accordance with Medical Devices Operator cleaning effect.
Generally a reduced suction capacity is a sign of possible 8.1.6 Device surface oversuction The ATMOS C 161 / C 261 Aspirator has a condensate collector at the bottom of the device. For visual • Prior to using the device on another patient, the complete inspection please remove the cap.
Cleaning- and servicing plan for ATMOS A / C 161 / 261 Aspirator Start date: Device name: Serial number: Exchange Cleaning/exchange performed Cleaning Cleaning Cleaning Exchange of of the DDS Exchange secretion canister the hous- suction hose bacterial and of fingertip...
Page 16
ATMOS partner or an ATMOS is half-full. Only use original ATMOS DDS bacterial and viral filters! authorized, specialized dealer. The suction device may not be operated without a DDS bacterial...
9.3 Sending in the device accessories or consumables. • Before passing on the ATMOS A / C 161 Aspirator, the 1. Remove all consumables and dispose of them properly. device must be brought into a technically and hygienically 2.
10.0 Troubleshooting The ATMOS A / C 161 / 261 Aspirator was subjected to a thorough quality control in the factory. If there is, nevertheless, some malfunction, you possibly might solve this problem yourself if you observe the following instructions.
Canister lid (incl. spare Canister lid DDS, blue, with spare DDS bacterial and viral filter 313.0006.0 DDS bacterial and viral Separate rinsing canister set for ATMOS A and C class incl. lid and 313.0008.0 filter) support Hose connector for hoses Ø 6 mm 000.0836.0...
(Used Electrical Appliances Register), this type of device is excluded from ElektroG regulations. In order to guarantee a proper disposal of your old device, please either pass on your old device to your specialised dealer or send it directly to ATMOS MedizinTechnik for a professional disposal.
IEC 61000-3-3 The ATMOS A / C 161 / 261 Aspirator may not be used directly next to or stacked with other devices. If operation close to or stacked with other devices is necessary, please watch the device to check its intended operation in this arrangement.
Page 24
The ATMOS A / C 161 / 261 Aspirator is designed for operation in the electromagnetic environment specified below. The customer or user of the ATMOS A / C 161 / 261 Aspirator should ensure that it is used in such an environment.
Page 25
The ATMOS A / C 161 / 261 Aspirator is designed for operation in the electromagnetic environment specified below. The customer or user of the ATMOS A / C 161 / 261 Aspirator should ensure that it is used in such an environment.
Page 26
The ATMOS A / C 161 / 261 Aspirator is intended for use in electromagnetic environment in which radiated disturbances are controlled. The customer or user of the ATMOS A / C 161 / 261 Aspirator can thereby help to prevent electromagnetic interference by maintaining a minimum distance between portable and mobile HF communication equipment (transmitters) and the ATMOS A / C 161 / 261 Aspirator - depending on the output of the communication device as indicated below.