6.3 Sending in the device

1. Remove and properly dispose of consumables.
2. Clean and disinfect the products and accessories according to the operating
instructions.
3. Place used accessories with the device.
4. Fill in the form QD 434 „Delivery complaint / return shipment" and the respective
Decontamination certificate.
 This form is enclosed to each delivery and can be found at www.atmosmed.com.
5. The device must be well padded and packed in suitable packaging.
6. Place the form QD 434 „Delivery complaint / return shipment" and the respective
decontamination certificate in an envelope.
7. Affix the envelope to the outside of the package.
8. Send the product to ATMOS or to your dealer.

6.4 Handling of batteries

Batteries are wearing parts and therefore excluded from the general warranty. There is
a function guarantee of 6 months.
Please observe the following notes in order to reach the maximum service life of your
battery:
Only use the original lithium-ionic battery 4IMR 19/66-2 BM18650Z3.
Please observe the operating instructions of the battery manufacturer.
Prior to first use the battery must be fully charged.
Battery-run devices should only be stored when they are charged.
Please fully recharge the battery every 6 months, even if the device is not used.
Prevent the batteries from direct solar radiation and keep them away from
radiators. The perfect storage temperature is between 8 and 15° C.
Exchange the battery when the remaining battery service life noticeably decreases.
Batteries are run-down after approx. 500 charging cycles.