Operation; Switch On The Device; Switch Off The Device; Vacuum Adjustment - Atmos E 341 Battery Operating Instructions Manual

Hide thumbs Also See for E 341 Battery:
Table of Contents

Advertisement

4.0 Operation

WARNING
Risk of infection by lack of hygiene or damaged components.
Deadly diseases can be transmitted.
Please use new consumables and new disposable canister systems or reprocessed
DDS canister systems for each patient.
Prior to each use, please check whether hoses or canister systems are damaged.
Replace any damaged parts.
Electric shock from damaged equipment.
Cardiac arrhythmias may be caused.
Prior to each use, please check whether the device and the recharging accessories
are damaged.
Replace any damaged parts immediately.
Do not use the device if it is damaged.
Ambient conditions during operation
Temperature:
Relative air humidity:
Air pressure:

4.1 Switch on the device

 The device should only be left on as long as you need it. This way you can increase
the battery life.
1. Push the on /off button to switch on the device.
The pump starts. The vacuum is set which was selected prior to switch off.
»
All LEDs on the control panel light up for about 1 second.
»
The on/off button is illuminated as long as the device is switched on.
»

4.2 Switch off the device

1. Switch off the device by pressing the on/off button for at least 1 second.

4.3 Vacuum adjustment

WARNING
Vacuum is too high.
Patient may be seriously injured.
Observe the valid guidelines.
Please select the vacuum according to the patient and the application.
1. Push the on /off button to switch on the device.
The pump starts. The vacuum is set which was selected prior to switch off.
»
2. Push the button for the required vacuum.
The green LED above the selected button flashes.
»
32
Operation
-5...+50° C
5...95 % without condensation
540...1100 hPa