4.2.10 DDS changeover docking station
4.2.11 Emptying the reusable canister system
NOTICE
The canister system must be cleaned and disinfected regularly. For more information,
see chapter "5 Reprocessing" on page 29.
NOTICE
Open the side drawer regularly to check the fill-level in the secretion canister.
 Empty the secretion canister manually when it is about half full.
 To do this, pull the secretion canister horizontally out of the support in the service
compartment and remove the suction hose.

4.3 Disposable canister system

4.3.1 Assembling the Receptal® canister system
Receptal® bag
Pump
Maximum load of the station: 15 kg!
• The DDS changeover docking station
for 2 canisters with changeover switch
1
and Direct Docking System may be
used with the ATMOS C 11 Systema.
• The changeover lever
switch the vacuum to the canister to
which the lever points.
• To remove or insert a canister, switch
the changeover lever towards the
canister that is not being replaced.
1. Insert the Receptal® suction bag into
the Receptal® external canister.
2. Close the canister tightly on all sides.
Patient
3. Check again for leaks; otherwise, no
vacuum can be built up.
4. Plug in the vacuum hose of the pump
and attach the secretion hose to the
connection marked "PATIENT".
 Only suction bags with an integrated
bacterial filter may be used.