10. Connect the suction hose.
11. Place the canister system upright from above into the support: Chapter „3.4
Connection and removal of canister system and hoses" on page 18.
12. Wrap the suction hose onto the hose rewind.
13. If you wish to use the device for vacuum mattresses: Check whether a suitable
adapter is available for the vacuum mattresses.

3.3 Charging the battery

 The battery status can be checked by briefly pressing the control button for the
battery status.
The battery must be fully charged prior to first use.
Damage to the battery due to deep discharge.
1. Charge the battery at the latest when the bottom green LED of the battery status
display flashes.
2. Only use the enclosed power supply and recharging unit 318.0035.0. Other charging
accessories must not be used.
3. Please observe the notes in chapter „6.4 Handling of batteries" on page 44.
During battery recharging full suction performance of the device is still available.
If the battery is fully discharged or defective, the device may be operated via the
charging accessories.
 If the ambient conditions are not adhered to, the charging time for the battery is
significantly increased. The charging process will be terminated if the temperature
is too high. Therefore, please prevent the device from direct solar radiation and
keep it away from radiators.
Ambient conditions during charging
Temperature:
Relative humidity:
Air pressure:
Charging with power supply and recharging unit
1. Connect the device plug from the power supply and recharging unit to the back of
the device
.
1
2. Connect the power cable to the power supply and recharging unit.
3. Plug in the power plug of the power supply and recharging unit to the socket.
+0...+40 °C
5...95 % without condensation
540...1100 hPa
1
Setting up and starting up
17