12.0 Notes on EMC

Medical electrical equipment is subject to special precautions with regard to EMC and must be installed
acc. to following EMC notes.
Portable and mobile HF communication facilities can influence medical electrical equipment.
The use of other accessories, other converters and cables than stated may lead to an increased emission
or a reduced interference immunity of the equipment or system.
Guidelines and Manufacturer´s Declaration - Emissions
The ATMOS
E 341 Battery is intended for use in the electromagnetic environment specified below. The
®
customer or user of the ATMOS
Emissions Test
RF Emissions acc.to CISPR 11
RF Emissions acc. to CISPR 11
Harmonic emissions according to
IEC 61000-3-2
Voltage fluctuations/flicker
according to IEC 61000-3-3
Guidelines and Manufacturer´s Declaration - Immunity for ATMOS
E 341 Battery
The ATMOS
E 341 Battery is intended for use in the electromagnetic environment specified below. The
®
customer or user of the ATMOS
IEC 60601- Test
Immunity Test
Level
ESD acc. to IEC
± 6 kV Contact
61000-4-2
± 8 kV Air
Fast electrical
± 2 kV Mains
transient/burst IEC
± 1 kV I/Os
61000-4-4
Surges IEC 61000-4-5
± 1 kV
common-mode
± 2 kV differential
mode
Voltage dips, short
< 5 % U
interruptions and
of the UT for 0.5
voltage variations
Cycle)
according to IEC
40 % U
61000-4-11
the UT for 5 Cycles)
70 % U
the UT for 25 Cycles)
< 5 % U
of the UT for 5 s)
Magnetic field at
3 A/m
power frequency
50/60 Hz acc. to IEC
61000-4-8
NOTE
U
is the AC mains voltage prior to application of the test level.
T
E 341 Battery should ensure that it is used in such an environment.
®
Compliance
Group 1
Class B
Class A
Corresponds
E 341 Battery should ensure that it is used in such an environment.
®
Compliance Level
± 6 kV Contact
± 8 kV Air
± 2 kV Mains
± 1 kV common-mode
± 2 kV differential
mode
(> 95 % Dip
< 5 % U
(> 95 % Dip
T
T
of the UT for 0.5
Cycle)
(60 % Dip of
40 % U
(60 % Dip of
T
T
the UT for 5 Cycles)
(30 % Dip of
70 % U
(30% Dip of
T
T
the UT for 25 Cycles)
(> 95 % Dip
< 5 % U
(> 95 % Dip
T
T
of the UT for 5 s)
Inapplicable
Electromagnetic Environment - Guidance
®
The ATMOS
E 341 Battery uses RF energy
only for its internal function. Therefore, its HF
emissions are very low and it is unlikely that
nearby electronic devices will be affected.
®
The ATMOS
E 341 Battery is suitable for use
in all establishments, including domestic,
and those directly connected to the public
low-voltage power supply network that
supplies buildings used for domestic purposes.
Electromagnetic Environment -
Guidance
Floors should be wood, concrete, or
ceramics tile. If floors are synthetic,
the relative humidity should be at
least 30 %.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
ATMOS
E 341 Battery demands
®
continued function even in case of
interruptions of the energy supply,
it is recommended to supply the
ATMOS
E 341 Battery from an
®
uninterruptible current supply or a
battery.
Power frequency magnetic
fields should be that of a
typical commercial or hospital
environment.
®
Notes on EMC
57