Maintenance And Cleaning - wtw SenTix 950 Operating Manual

Ph combination electrodes with refillable liquid reference system

Advertisement

SenTix
Refilling the
reference
electrolyte
ba75843e01
08/2009
Note
During longer storing periods, salt sediments may develop on the watering
cap. They do not affect the measuring characteristics and can easily be re-
moved with deionized water when the electrode is put into operation again.
Aging
Every pH electrode undergoes a natural aging process. With aging, the re-
sponding behavior becomes slower and the electrode slope and asymmetry
change. Moreover, extreme operating conditions can considerably shorten
the lifetime of the electrode. These are:
Strong acids or lyes, hydrofluoric acid, organic solvents, oils, fats, bro-
mides, sulfides, iodides, proteins
High temperatures
High changes in pH and temperature.
The warranty does not cover failure caused by measuring conditions and
mechanical damage.

Maintenance and cleaning

During operation, a small amount of reference electrolyte leaks through the
junction from the electrode into the test sample. If the level of reference elec-
trolyte becomes too low with time, refill it through the refilling opening. Refill-
ing is very easy with the enclosed dropping bottle. Proceed as follows:
Cut off the tip of the dropping bottle at a right angle until the opening in the
tip can be seen
Open the refilling opening of the electrode
Press the tip of the dropping bottle into the refilling opening while turning
it slightly
Pump the reference electrolyte in the shaft using the dropping bottle
Pull the dropping bottle out of the refilling opening while turning it slightly
as necessary.
Aging
17