Assembly And Installation; Scope Of Delivery - ProMinent DULCOMARIN 3 Compact Assembly And Operating Instructions Manual

Multi-channel measuring and control system
Table of Contents

Advertisement

6

Assembly and installation

6.1 Scope of delivery

Tab. 3: The following components are included as standard:
Description
Measuring/control system DULCOMARIN
Assembly material, complete, 2P universal (set).
Threaded cable connector set, M20 DCPa.
Data DVD.
Operating instructions.
Access card with PUK.
User qualification, mechanical installation: trained and qualified
n
Ä Chapter 1.5 'User qualification' on page 10
personnel
User qualification, electrical installation: electrical technician
n
Ä Chapter 1.5 'User qualification' on page 10
NOTICE!
Installation site and ambient conditions
The controller meets the requirements for IP 67
(housing, closed) or IP 20 (housing, open) and
leak-tightness (based on NEMA 4X). These
standards are only met if all seals and
threaded connectors are correctly fitted.
Only carry out the (electrical) installation after
(mechanical) installation.
Ensure that there is unimpeded access for
operation.
Ensure safe and low-vibration fixing.
No direct sunlight.
Permissible ambient temperature of the con‐
troller at the installation location: - 5 ... 50 °C at
max. 95% relative air humidity (non-con‐
densing).
Requires a low voltage cable with a tempera‐
ture resistance of ≧ 70 °C.
Take into consideration the permissible
ambient temperature of the connected sensors
and other components.
The controller is only suitable for operation in
closed rooms. If operating outdoors, use a suit‐
able protective enclosure to protect the con‐
troller from the environment.
Mounting position
As standard, a wall-mounted controller is used.
Always install the controller so that the cable
entries point downwards.
Leave sufficient free space for the cables.
®
3 including wall bracket.
Assembly and installation
Quantity
1
2
1
1
1
1