Calibrate: measured variable, ORP

Once the packaging has been opened, the ORP buffers often veer
towards higher mV values due to the addition of oxygen. Therefore
use as small a vessel as possible and use up the ORP buffer
quickly.
Please note the following general rules:
Never calibrate the sensor in the storage bottle, instead always
n
take out the amount of buffer needed and dispose of the
removed buffer after the calibration.
Never pour the buffer back into the storage bottle, but dispose
n
of it.
[CAL]
1.
Tap on
Calibration
Calibrate the measured value
pH
ORP
Temperature
Free chlorine
Total chlorine
Temperature of Cl sensor
Fig. 47: Calibration interface.
2.
Select the measured variable to be calibrated by tapping and
scrolling on the display.
3.
Follow the instructions on the display.