5.5.5

Electrical connection

A
L
N
+
-
+
52
1+ 2-
31 32
41 42 51
PE
Fig. 36: Electrical connections
A Connections for power supply and outputs B Connections for signal cable (remote mount design only)
Change from one to two columns
NOTE
For detailed information about earthing the transmitter and
the sensor, please refer to chapter "Grounding the flowmeter
sensor" on page 19!
Connections for the power supply
AC power supply
Terminal
Function / comments
L
Phase
N
Neutral conductor
PE /
Protective earth (PE)
DC voltage supply
Terminal
Function / comments
1+
+
2-
-
PE /
Protective earth (PE)
B
-
+
3
FE
2S E2 E1 1S
FE
3
2S E2 E1 1S
B
M1
M2
PE
M1
M2
PE
Connections for outputs
Terminal
Function / comments
31 / 32
Active current output
The current output is "active" mode. The source to drive
the 20 mA loop is in-built in the transmitter.
41 / 42
Passive digital output DO1
The output can be configured as a pulse output,
frequency output or switch output on site.
51 / 52
Passive digital output DO2
The output can be configured as a pulse output,
frequency output or switch output on site.
Functional earth
Connections for the signal cable
Only for remote mount design.
Terminal
Function / comments
FE
Not connected
3
Measurement potential
2S
Shield for E2
E2
Signal line
E1
Signal line
1S
Shield for E1
M1
Magnet coil
M2
Magnet coil
Shield
Not connected
ProcessMaster FEP610, HygienicMaster FEH610 | OI/FEP610/FEH610-EN Rev. B 25
PE
≤ 50 m (200 m)
≤ 164 ft (656 ft)
PE
G12018
Color
green
blue
violet
brown
red
orange / yellow