Error no. /
Error text
Range
M038.010 /
Sensor memory defective.
Electronics
Mem. or connect. defective.
Replace memory.
M037.015 /
NV chips defect on
Electronics
Motherboard.
Defective MB. Replace MB.
Call Service.
M032.023 /
Curr.Out 31 / 32 not
Config.
calibrated.
Call Service.
M028.008 /
Display value is < 1600 h at
Config.
Qmax.
Change mass Unit or vol.
Unit for Totalizer.
M026.005 /
Maintenance interval is
Config.
reached.
Perform maintenance.
M024.009 /
Device not calibrated.
Config.
Call Service.
Change from one to two columns
70 OI/FEP610/FEH610-EN Rev. B | ProcessMaster FEP610, HygienicMaster FEH610
Current output
Digital output
Current value - no
No response.
change.
High Alarm or Low
Current value - no
Alarm, depending
change.
on parameter
"Curr.Out at
Alarm".
Pulse output
LCD display
Current value - no
Current value - no
change.
change.
Current value - no
Current value - no
change.
change.
Error maskable?
Menu "...Group
Masking".
Menu "...Group