Menu: Diagnostics

Menu / parameter
Diagnostics
...Maintenance control
...Diagnosis Control
...Diagnosis Values
......Simulation Mode
...Output Readings
Diagnostics / ...Maintenance control
Preset Maint. cycle
Maint. Remain. Time
Start New Cycle
Diagnostics / ...Diagnosis Control
...Empty Pipe Detector
...Noise Check
...Fingerprints
Diagnosis Interval
Diagnostics / ...Diagnosis Control / ...Empty Pipe Detector
Empty Pipe Detector
Adjust EP
Manual Adjust EP F.
Threshold
Detector EP Value
Description
Selection of submenu "...Maintenance control" using
Selection of submenu "...Diagnosis Control" using
Selection of submenu "...Diagnosis Values" using
Selection of submenu "......Simulation Mode" using
Selection of submenu "...Output Readings" using
Sets the service interval (0 ... 9999 hours).
After the service interval has expired, the corresponding error message "Maintenance interval is reached.
Perform maintenance." is set. The setting "0" deactivates the maintenance interval.
Time remaining in the maintenance interval until the error message "Maintenance interval is reached.
Perform maintenance." is set.
Resetting of the maintenance interval. The service interval is reset to the value set in "Preset Maint. cycle".
Selection of submenu "...Empty Pipe Detector" using
Selection of submenu "...Noise Check" using
Selection of submenu "...Fingerprints" using
Sets the time span between performing the individual diagnostic.
The "Empty Pipe Detector" function will be performed in the selected time span (e. g. every 5 seconds).
Default setting: 5s.
Activate the "Empty Pipe Detector" function (only for sizes ≥ DN 10).
An entirely full measuring tube is essential for an accurate measurement. The "Empty Pipe Detector" function
detects an empty measuring pipe.
In the case of an alarm, the current output assumes the status that was defined in the "Input/Output / ...Curr.Out
31 / 32 / Curr. at EPD Alarm" menu, and the pulse output is stopped.
The Empty Pipe Detector function must be adjusted according to the conditions on site. The switching threshold is
set during the automatic adjustment.
Start automatic adjustment of the Empty Pipe Detector function.
Manually adjust the Empty Pipe Detector function.
The value must be modified in such a way that the frequency for empty pipe detection (Detector EP Value) is close
to 2000 Hz.
NOTE
Prior to starting the (manual / automatic) adjustment, make sure that:
— There is no flow through the flowmeter sensor (close all valves, shut-off devices, etc.)
— The flowmeter sensor is completely filled with the fluid to be measured.
Set the switching threshold for empty pipe detection.
The switching threshold is set automatically during automatic adjustment. The switching threshold can be changed
in order to obtain manual fine adjustment.
Display of the frequency for empty pipe detection. If the current value exceeds the defined switching threshold, a
message appears on the display and an alarm is output via the digital output, if configured accordingly.
.
.
.
.
.
.
.
.
ProcessMaster FEP610, HygienicMaster FEH610 | OI/FEP610/FEH610-EN Rev. B 61