Menu / parameter
Device Setup / ...Transmitter / ...Units
Unit Massflow Qm
Unit Mass Tot.
Unit Volumeflow Qv
Unit Vol. Totalizer
Cust Qm Tot Unit F.
Cust Qm Tot Unit N.
Cust Qv Tot Unit F.
Cust Qv Unit N.
Device Setup / ...Transmitter / ...Cut Off
Low Flow Cut Off
Low Flow Hysteresis
Device Setup / ...Transmitter / ...Feature Settings
Verifying Capab.
Verifying Capab.Code
Device Setup / ...System Zero
Manual Adjust
Auto Adjust
Description
Selection of unit for mass flow.
Refer to "Available units" on page 49.
The selection applies to the display of the current mass flow, and for the parameters related to mass flow such as
QmMax and Qm
DN.
Max
Selection of unit for the mass totalizer.
Refer to "Table 3: Units for the mass totalizer" on page 49.
Selection of unit for volume flow.
Refer to "Table 1: Units for the volume flow" on page 49.
The selection applies to the display of the current volume flow and for the parameters related to volume flow such
as QvMax and Qv
DN.
Max
Selection of unit for the volume totalizers.
Refer to "Table 4: Units for the volume totalizer" on page 49.
Enter the factor for a user-defined massflow unit. The factor relates to the flow per liter.
Enter the name for the user-defined massflow unit.
Enter the factor for a user-defined volumeflow unit.
The factor relates to the flow per liter.
Enter the name for the user-defined volumeflow unit.
Sets the switching threshold (0 ... 10 %) for the low flow cut-off.
If the flow rate is below the switching threshold, there is no flow measurement. The setting of 0 % deactivates the
low flow cut-off.
Default setting: 1.0 %
Sets the hysteresis (0 ... 50 %) for the low flow cut-off as it is defined in the parameter "Low Flow Cut Off".
Default setting: 20 %
Indicates whether the Verification function is enabled.
Sets the device-specific code for activating the Verification function. To use this function subsequently, contact the
ABB service team or sales organization.
After entering the code, restart the device (e.g. using the parameter "Device Reset" or by briefly switching off the
power supply).
Sets the value for zero point adjustment in % of Q
Manual Adjust: -50 ... +50 mm/s
Starts the automatic zero point adjustment using
NOTE
Prior to starting the zero point adjustment, make sure that:
— There is no flow through the sensor (close valves, shut-off devices etc.).
— The sensor must be filled completely with measuring medium for measurement.
DN.
max
. Automatic zero point adjustment takes approx. 60 seconds.
ProcessMaster FEP610, HygienicMaster FEH610 | OI/FEP610/FEH610-EN Rev. B 55