Electrical Connection Area; Installations For Protection Class Ip 68 - ABB COPA-XE Instruction Bulletin

Electromagnetic flowmeter with pulsed dc magnetic field excitation
Table of Contents

Advertisement

Electromagnetic Flowmeter
COPA-XE / MAG-XE
9.2.3

Electrical Connection Area

Supply Power{
In-/Outputs
Ground
Signal/Excitation Cable
Fig. 31
Converter, Field Mount Housing
View into Connection Box
Note:
!
When installing the signal/excitation cable at the
flowmeter primary a water trap should be provided
(Fig. 32).
Fig. 32
Cable Installation with a Water Trap
44
SE-Terminal
Water Trap
9.2.4

Installations for Protection Class IP 68

The max. submergence depth for flowmeter priamries designed
for Protetion Class IP 68 is 5 m. A hose enclosed connector is
used in place of the standard PG connector. The signal/
excitation cable must be enclosed in the 1/2" hose from the
connection box to above the maximum submergence level
(Fig. 33). Above this level the cable is installed using the cable
connector included with the shipment in a water tight manner.
The hose is then attached to the hose connector using the
threaded clamp ring. Finally the connection box should be
carefully secured.
Signal/Excitation Cable
Water Tight Cable Connector
Install the hose above the submergence
level and seal the signal/excitation cable
using the cable connector.
Fig. 33
Installation IP 68 (Hose Connector)