Hygienicmaster - ABB ProcessMaster FEP611 Operating Instructions Manual

Table of Contents

Advertisement

2.1.2

HygienicMaster

Integral mount design
1
Fig. 2:
Designs
1 Flowmeter sensor 2 Remote transmitter
Flowmeter sensor
Model
Housing
Measuring accuracy for liquids
Permissible measuring medium
temperature T
medium
Minimum conductivity
Nominal pressure
Nominal diameter
Process connection
Process connection material
Liner material
Electrode material
IP rating
Approvals
Pressure Equipment Directive
2014/68/EU
CRN (Canadian Reg.Number)
Hygienic design approvals
Further approvals
1) 3A approved
FEH611
HygienicMaster FEH611, FEH612
Integral mount design, remote mount design
0.5 % of measured value
-25 ... 130 °C (-13 ... 266 °F)
> 20 μS/cm (20 μS/cm for demineralized water)
PN 10 ... 40, ASME CL 150, 300, JIS 10K
DN 3 ... 100 (1/10 " ... 4 ")
Wafer type design:
Flange in acc. with DIN, ASME or JIS:
Threaded pipe connection in acc. with DIN 11851
1)
Weld stubs
:
Tri-Clamp in acc. with DIN 32676
Tri-Clamp in acc. with ASME BPE
External thread in acc. with ISO 228 / DIN 2999:
Stainless steel
PFA (vakuum tight)
CrNi steel 1.4571 (AISI 316Ti), 1.4539 [904L], Hastelloy B, Hastelloy C, platinum-iridium, tantalum, titanium
Integral mount design: IP 65 / IP 67
Remote mount design: IP 65 / IP 67 / IP 68 (sensor only)
Conformity assessment in accordance with category III, fluid group 1
On request
3A, FDA-approved materials
At www.abb.com/flow or on request.
1
FEH612
DN 3 ... 100 (1/10 ... 4")
DN 3 ... 100 (1/10 ... 4"), PN 10 ... 40
1)
:
DN 3 ... 100 (1/10 ... 4"), PN 10 ... 40
DN 3 ... 100 (1/10 ... 4"), PN 10 ... 40
1)
:
DN 3 ... 100 (1/10 ... 4"), PN 10 ... 16
1)
:
DN 3 ... 100 (1/10 ... 4"), PN 10
DN 3 ... 25 (1/10 ... 1"), PN 16
ProcessMaster FEP610, HygienicMaster FEH610 | OI/FEP610/FEH610-EN Rev. B 7
Remote mount design
2
FET612
G12002

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents