Connection to remote mount design

NOTE
Use wire end sleeves.
— Wire end sleeves 0.75 mm
(1S, 2S).
— Wire end sleeves 0.5 mm
The shields may not touch (signal short circuit).
Flowmeter sensor site
20
(0.8)
9
8
70 (2.76)
25 (
Fig. 40: Signal Cable Part No. D173D031U01, Dimensions in
mm (inch)
Pos.
Terminal
1
3
2
E1
3
1S
4
E2
5
2S
6
M2
7
M1
8
9
SE /
Change from two to one column
28 OI/FEP610/FEH610-EN Rev. B | ProcessMaster FEP610, HygienicMaster FEH610
2
(AWG 19), for shielding
2
(AWG 20), for all other wires.
90 (3.54)
6
A
7
A
8 (
0.31
)
0.98
)
Function / comments
Measurement potential
Signal line
Shield for E1
Signal line
Shield for E2
Magnet coil
Magnet coil
Not connected
Not connected
Shield
Transmitter side
20
(0.8)
1
2
3
4
5
Fig. 41: Signal Cable Part No. D173D031U01, Dimensions in
mm (inch)
Pos.
Terminal
1
3
2
E1
3
1S
4
E2
5
2S
G12025
6
M2
7
M1
8
Color
9
SE /
green
violet
blue
red
brown
yellow
orange
70 (2.76)
50 (1.97)
A
8 (
0.31
25 (
0.98
)
Function / comments
Measurement potential
Signal line
Shield for E1
Signal line
Shield for E2
Magnet coil
Magnet coil
Not connected
Not connected
Shield
1
2
3
4
5
A
)
G12026
Color
green
violet
blue
red
brown
yellow
orange