Operating The Dmt; Permanent Display; Probe Condition Information (Ph Only) - ProMinent DULCOMETER DMT Series Operating Instructions Manual

On-site measurement transducer measured variables ph/redox/temperature
Hide thumbs Also See for DULCOMETER DMT Series:
Table of Contents

Advertisement

Operating the DMT

8
8.1
pH
12.5 mA
Redox
12.5 mA
The permanent display allows you to monitor the DMT measured values
during operation.
The permanent display can indicate the following (depending upon
general settings):
8.1.1
Zero point:
Slope:
Pre-set time:
Page 38
Operating the DMT

Permanent display

7.12
pH
Information about the probe condition
21.0 ˚C
312
mV
21.0 ˚C
Measured variable
Correction variable
Output current

Probe condition information (pH only)

Display:
Zero point
Slope
Pre-set time
Admissible range
Optimum range
-60
-50
0
+50
Admissible range
Optimum range
40
45
59,16
63
0
30
Information about the probe condition
Error messages
Condition:
visible = probe OK
not visible = on the limit of admissible range
and/or not in order
m
+60
V
mV/pH
65
The pre-set time depends upon the
temperature and the cable length.
s
Depending on the system the setting
time can be more than 60 seconds.