Parameter Output - ProMinent DULCOMETER DMT Series Operating Instructions Manual

On-site measurement transducer measured variables ph/redox/temperature
Hide thumbs Also See for DULCOMETER DMT Series:
Table of Contents

Advertisement

DMT settings
7.1.5.5 Calibrating the PT 100 temperature gauge

7.1.6 Parameter output

Depending upon the "output" general setting you have the following settings
options:
7.1.6.1 General setting: output current = measured value
During calibration or checking of the ORP probe, the output current increases
to 23 mA.
Page 28
NOTE
You need only calibrate the temperature gauge if
- You have a PT 100 temperature gauge and the sensor cable
is longer than 4 m.
- You have a precision measurement device (the DMT
measures to ±0.5 °C / ±0.9 °F accuracy).
• Do not replace the temperature sensor during calibration!
Immerse the DMT temperature gauge and the reference measurement
device into the same liquid sample.
Read off the value from the reference measurement device once the
temperature is stable.
Set the correction value under "offset" until the temperature value is
identical to the reference value.
(= 0 in identity code)
Under "4 mA" enter the value at which the output current should be
4 mA.
Under "20 mA" enter the value at which the output current should be
20 mA.