ProMinent DULCOMETER DMT Series Operating Instructions Manual page 27

On-site measurement transducer measured variables ph/redox/temperature
Hide thumbs Also See for DULCOMETER DMT Series:
Table of Contents

Advertisement

7.1.5.2 General setting: Temperature measurement = manual
pH
param. temperature
input: manual
unit: ˚C
7.1.5.3 General setting: Temperature measurement = automatic or
pH
param. temperature
input: manual
unit: ˚C
The selection made under "input" affects subsequent settings options.
Sensor
If you have selected sensor in the "input" option:
The actual temperature measured is displayed under "meas. val.".
Manual
If you have selected manual in the "input" option:
7.1.5.4 General setting: Temperature measurement = none
No info display appears in the operating menu.
No temperature measurement is carried out.
ProMinent
®
(pH only)
Enter the process temperature under "T".
Select the temperature unit °C or °F under "unit".
manual (pH only)
Under "input" select sensor or manual:
Select sensor if a temperature gauge is connected (temperature
measurement = automatic):
Select manual if the user is going to enter the process temperature
(temperature measurement = manual).
Select the unit of measurement for temperature °C or °F under "unit".
Enter the difference ∆t under "offset" to a reference temperature
measurement (see 7.1.5.5 Calibrating the Pt 100 temperature gauge).
Enter the process temperature under "T".
Select the temperature unit °C or °F under "unit".
T = 22.6 ˚C
unit: ˚C
input: sensor
unit: ˚C
offset = 0.00 ˚C
meas. val. = 21.00 ˚C
or
input: manual
T = 22.6 ˚C
Unit: ˚C
DMT settings
Page 27