The Calibration Process - ProMinent DULCOMETER DMT Series Operating Instructions Manual

On-site measurement transducer measured variables ph/redox/temperature
Hide thumbs Also See for DULCOMETER DMT Series:
Table of Contents

Advertisement

Operating the DMT
8.2.2
Calibration menu overview
8.2.2.1 2 point calibration
Page 40

The calibration process

NOTE
• The back key
any time. You will then return to the permanent display.
• The output current is frozen for the duration of the calibra-
tion at, depending upon general settings, the last
measured value (HOLD) or at 23 mA (see chapter 7.1.7
General settings).
• Inadmissible values render the ongoing calibration invalid.
The previous calibration values are retained. The error
messages are explained in chapter 9.2.
• Please dispose of spent buffer solution.
can be used to exit the calibration menu at
ProMinent
®