7.1.7 General settings

pH Redox
general settings
DMTaW090P10E0220
7.1.7.1 Changing the device type
You can select the measured variables, pH or redox potential, in the first
menu option "device type".
Depending upon your choice the corresponding identity code position alters:
pH = "P", redox = "R".
If you have confirmed the change of measured variable with the enter key
the output current will change immediately.
The settings for each measured variable are unaffected even after a different
selection has been made and are reinstated if the selection is reversed.
7.1.7.2 Changing general settings
You can adapt the DMT to your individual process requirements in the
second menu option "general settings".
Access the individual identity code positions using the change key
Non-alterable features are automatically bypassed.
The following tables show the settings options depending upon the selected
measured value (the sequence from left to right corresponds to the identity
code position).
Measured variable pH
Language
Automatic buffer
detection
D = German
0 = ProMinent
E = English
F = French
D = DIN 19266
P = Polish
I =
Italian
V = automatic
S = Spanish
ProMinent
®
2.1.06
Under "type" select pH or ORP (redox).
Temperature
measurement
0 = automatic
buffer set
1 = manual
2 = automatic or
buffer set
manual
9 = none
adjustment
device type
!! correct probe ??
type = pH
DMTaW090P10E0220
general settings
change
access code = 5000
free access
Output
0 = measured variable;
fault 23 mA,
at calibration 23 mA
1 = manually adjustable
current value
2 = measured variable
or manual:
at fault 23 mA,
at calibration 23 mA
3 = measured variable
or manual:
at fault 23 mA,
when calibrating the last
measured value (HOLD)
4 = constant 4 mA
current
DMT settings
general settings
language
DMTaW090P10E0220
E = english
.
Other
settings
0 = standard
Page 31