Assembly and installation
4.2.1
Fig. 5
Then follow the steps below:
Page 14

Wall mounting

Unfasten the four housing screws
Lift the front section slightly forwards and then swing open to the left.
Pierce threaded cable apertures at the bottom of the rear side as required
(fig. 5).
NOTE
• The large gland (M 20 x 1.5) is intended for the sensor
cable.
• Feed the power cable leftwards through the smaller
threaded connector (M 16 x 1.5). Follow with the other
sensor cables (e.g. PT 1000).
Tighten the threaded connectors (fig. 6, ➀) as required.
Use reducers as required (fig. 6, ➁) to adapt the sizes of the threaded
connectors to the actual cable diameters.
Feed the cables into the threaded connectors.
Now proceed according to 4.2.4 Connecting coaxial cable and
4.2.5 Connecting terminals.
Tighten the locking screws (fig. 6, ➂) for the threaded connectors.
Swing the front section onto the back section.