7.1

7.1.1 Menu overview: measured variable pH

pH
12.5 mA
param. measurement
zero point. = -0.1 mV
slope = 60.30 mV/pH
probe check: on
param. temperature
input: sensor
unit: ˚C
param. output signal
output: meas. value
4 mA: 2.00 pH
20 mA: 12.00 pH
general settings
DMTaW090P10E0220
ProMinent
®

Measured variables pH and redox

NOTE
The following menu is only an example.
The displays can vary depending upon the general settings.
7.12
pH
21.0 ˚C
2.1.06
= Request access code
zero point = -0.1 mV
slope = 60.30 mV/pH
probe check: on
meas. val. = 7.12 pH
input: sensor
unit: ˚C
offset = 0.00 ˚C
meas. val. = 21.00 ˚C
output: meas. value
4 mA: 2.00 pH
20 mA: 12.00 pH
device type
!! correct probe ??
type = pH
DMTaW090P10E0220
general settings
change
access code = 5000
free access
DMT settings
(if activated by the user)
general settings
language
DMTaW090P10E0220
E = english
Page 23