Assembly and installation
Pipe mounting, please follow the steps below:
4.1.3
Fig. 4
Page 12
Remove wall/pipe bracket from DMT:
pull the two snap fasteners outwards and push upwards (fig. 1, ➀).
Swing the wall/pipe bracket away from the DMT and pull downwards
(fig. 1, ➁).
Fasten the bracket to the pipe using the cable ties or pipe clamps (fig. 3).
If the DMT is also to be secured with a screw, pierce the screw hole in the
back of the housing (housing must be open) and attach a rubber insert
(fig. 2, ➀) to the bracket.
Hang the DMT onto the top of the bracket (fig. 3, ➀) and push down
gently against the bracket (fig. 3, ➁), then push upwards until you hear a
click (fig. 3, ➂).
Tighten the locking screw to secure more firmly (housing is open).

Panel mounting

IMPORTANT
• The control panel must be thick enough to withstand the
weight of the mounted DMT without buckling. (To achieve
enclosure rating IP 54, steel must be at least 2 mm thick;
plastic should be correspondingly thicker).
NOTE
The DMT will protrude from the control panel approx. 30 mm
once mounted.