calibration
The calibration values (zero point and slope) are now displayed:
pH
zero point = -0.12 mV
slope = 58.32 mV/pH
meas. val. = 7.21 pH
The current pH value is shown under "meas. val.".
The display changes to permanent display. The calibration is complete.
ProMinent
®
After completion of this calibration stage the display shows the next
menu option for 2 point calibration:
The buffer value can be corrected once again at this point using the arrow
keys
/
.
Press the enter key
Proceed as for "conclude calibration".
acceptance
Press the enter key
without saving the values.
NOTE
Inadmissible values render the current calibration process
invalid. The previous calibration values are retained.
The error messages during calibration are explained in
chapter 9.2.
to end 2 point calibration.
to adopt the values or the back key
Operating the DMT
to exit