DMT settings

7.2.2 Parameters temperature

The actual measured temperature is displayed under "meas. val.".

7.2.3 Parameter output

You have the following settings options, depending on the "output" type you
selected under general settings:
7.2.3.1 General setting: output current = measured variable
7.2.3.2 General setting: output current = adjustable
This setting is used for the function check.
Page 34
Select the temperature unit ˚C or ˚F under "unit".
Enter under "offset" the difference ∆t from a reference temperature
measurement (see 7.1.5.5 Calibrating the Pt 100 temperature gauge).
NOTE
• The DMT automatically recognises whether a PT 1000 or a
PT 100 has been connected.
(= 0 in the identity code)
Under "4 mA", enter the value at which the output current should be 4 mA.
Under "20 mA", enter the value at which the output current should be
20 mA.
(= 1 in the identity code)
Under "l", enter an output current between 4 and 20 mA.