ProMinent DULCOMETER DMT Series Operating Instructions Manual page 32

On-site measurement transducer measured variables ph/redox/temperature
Hide thumbs Also See for DULCOMETER DMT Series:
Table of Contents

Advertisement

DMT settings
Measured variable redox
Language
Probe type
D = German
0 = standard
E = English
F = French
P = Polish
I = Italian
S = Spanish
7.1.7.3 Changing access code
The factory pre-set access code is 5000. This setting allows unlimited access
to the calibration menu and the menu options in the settings level.
This code can be changed in the third menu option "change access code".
Use the change key
The new code is confirmed and activated with the enter key
Page 32
Temperature
measurement
0 = automatic
9 = none
to move to individual digits.
NOTE
• If an access code has been activated, both the calibration
menu and the menu options are access-protected.
To gain access, enter the access code when requested
to do so and confirm using the enter key
protection is reactivated when the DMT returns to
continuous display mode.
• Deactivate the access code by resetting the code to 5000
in the "change access code" menu option.
Output
0 = measured variable:
at fault 23 mA,
at calibration 23 mA
1 = manually adjustable
current value
2 = measured variable
or manual:
at fault 23 mA,
at calibration 23 mA
3 = measured variable
or manual:
at fault 23 mA,
when calibrating the last
measured value (HOLD)
4 = constant 4 mA
current
Other
settings
0 = standard
.
. Access