IMPORTANT
• The signal inputs of all devices in the current loop must be
electrically isolated from the current output.
• Take into account the sum of the ohmic resistances of all
devices in the current loop (excluding power supply).
The input voltage of the transducer must never drop below
16 V during operation.
The measured value will otherwise be wrong.

Wiring example - two wire system