Create Configuration File 4 - Single Clock with Marker

1
2
3
4
5
6
7
Keysight 16860 Series Portable Logic Analyzer Service Guide
In the Save As dialog, type the file name SingleClock_NoMarker.ala. For file options, select Setup
Only. Then click Save.
Start with the MultipleClocks_WithMarker configuration file. Select File and then Open. In the Open
dialog select the file then open. If asked about saving the current configuration, click No.
Ensure that there is a U4203A connected to Pods 1 and 2. If Pods 3 and 4 exist, connect a
U4203A to these pods. The Pod 1 leads should be connected to the generator as described
above.
Open the Analyzer Setup Sampling tab.
In the State Options section, change the Sampling options to Single Clock.
In the clock selection area, change the Pod 1 clock to Both Edges.
Close the dialog by clicking OK.
Save the configuration by selecting File then Save As ....
Testing 16860 Performance