Removing and Replacing the Acquisition Tray
Ensure that you have:
•
•
•
To remove and replace the acquisition tray of 16860 logic analyzer
1
2
3
Keysight 16860 Series Portable Logic Analyzer Service Guide
completed the tasks listed in the
removed the cover using the procedure listed in
Acquisition Bezel, and Cover"
removed the front panel using the procedure listed in
Panel"
on page 112.
Disconnect the four (Power, SATA, and flat) cables from the acquisition board. These cables are
pointed out in the following photo.
For reconnection later, note the relative locations of red and blue SATA
NOTE
cables.
Unthread the cables back through the grommeted hole.
Using a Torx T10 screwdriver, remove the following screws securing the tray assembly to the
chassis:
Removing, Replacing, or Returning 16860 Series Logic Analyzer Assemblies
"Before you Start
topic.
"Removing and Replacing the Hard Drive,
on page 109.
"Removing and Replacing the Front
6
119