Keysight 16860 Series Portable Logic Analyzers Service Guide
Disable all analyzers except the one being tested. This simplifies the instructions and makes
module initialization faster.
a Select the Overview tab at the bottom of the main window.
Set up the bus:
a In the Overview window, select Setup -> Bus/Signal... from the module's drop-down menu.
b Scroll if necessary to view the pods you are testing.
c Verify that the pod threshold buttons say "Threshold: Differential". If they don't, make sure the
correct probes (E5379A) are attached to pods 1 and 2. The threshold is set to Differential
automatically when E5379A probes are attached.