Single Clock Falling Edge - Maximum Clock Rate

1
2
3
4
5
6
7
8
Keysight 16860 Series Portable Logic Analyzer Service Guide
Load the SingleClock_NoMarker.ala configuration file.If asked to save the current configuration file,
click No.
Verify that the Generator is set to 714 MHz as a starting point.
Open the Sampling tab in the Analyzer Setup dialog by clicking the Sampling Setup Icon.
In the clock assignment area, set Pod 1 Clock to Falling Edge.
Close the dialog windows by clicking OK.
Click the Run Repetitive
acquiring data repeatedly. Acquired data will start appearing in the Listing window.
Start increasing the frequency on the pulse generator by 1 MHz increments while simultaneously
observing the logic analyzer data acquisition status.
Continue increasing the generator frequency until the logic analyzer displays an error that the
data could not be displayed, or the clock is too fast. It is possible that the error is caused by noise
introduced as the generator frequency is changed.
toolbar button to start a repetitive run on the logic analyzer for
Testing 16860 Performance