d Channels 7 through 0 are already assigned by default. Assign pod 2 channels 15 through 0 and
pod 1 channels 15 through 8 by clicking and dragging from the left-most channel box to the
right-most channel box. Your display should look like the lower picture when you are done.
Select the State sampling mode and set the State Clock options:
a Select the Sampling tab of the Analyzer Setup window.
b Select State - Synchronous Sampling.
c For State Clock, select CLK1 Clock and Both edges.
Set the trigger position and acquisition memory depth:
a Set the Trigger Position to 100% Poststore.
b Set the Acquisition Depth to 8K.