Single Clock Falling Edge - Pod 1 Data - Setup For Maximum Data Rate - Keysight Technologies 16861A Service Manual

Portable logic analyzers
Table of Contents

Advertisement

3
Testing 16860 Performance
9
10 If the analyzer displays an error again, lower the frequency by 1MHz and run again. If the test
11 Continue changing the frequency and running unit the frequency can no longer be increased and
12 Set the generator to the highest frequency that did not cause an error.
13 Click Run Repetitive
14 Wait for logic analyzer to complete 100 acquisitions at the new pulse generator frequency
15 Click the Stop
16 Note the generator frequency setting. This will be used in the next section to verify the data rate.

Single Clock Falling Edge - Pod 1 Data - Setup for Maximum Data Rate

1
2
3
68
Without changing the frequency, clear the message by clicking OK. Then click the Run Repetitive
toolbar button to start a repetitive run again at the same frequency. Running at the same
frequency checks to see if the error message was caused by noise due to the changing of the
generator frequency.
does not display an error, increase the frequency by 1 MHz and run again.
the error message not displayed.
.
without displaying any error. If an error is displayed, decrease the pulse generator frequency by 1
MHz and then again wait for 100 acquisitions at this new frequency without any error. Repeat this
step until you get 100 acquisitions without any error display.
toolbar button to stop the data acquisition.
On the 81134A, verify that Channel 2 Freq Divide is set to 2.
Load the SingleClock_WithMarker.ala configuration file. If asked to save the current configuration
file, click No.
Verify that the Generator is set to the frequency found in the last section as a starting point.
Keysight 16860 Series Portable Logic Analyzer Service Guide