Keysight 16860 Series Portable Logic Analyzer Service Guide
Remove the touchscreen cable from the front panel. To do this, first unclip the three clips that
hold this cable to the front panel. These clips are pointed out in the photo below. Then disconnect
the cable from the two locations encircled in red in the photo below.
Disconnect the touchscreen flat cable from the USB Controller Board as pointed out in the photo
below.
For reinstallation later, note that the touchscreen flat cable is connected
NOTE
to the lower pins on the USB Controller Board as pointed out in the photo
below.
Removing, Replacing, or Returning 16860 Series Logic Analyzer Assemblies