Testing 16860 Performance
Create Configuration File 3 - Single Clock (no Marker)
1
2
3
4
5
6
7
42
Start with the MultipleClocks_NoMarker configuration file. Select File then Open. In the Open dialog
select the file and then open. If asked about saving the current configuration, click No.
Ensure that there is a U4203A connected to Pods 1 and 2. If Pods 3 and 4 exist, connect a
U4203A to these pods also. The Pod 1 leads should be connected to the generator as described
above.
Click the Analyzer Setup > Sampling tab.
In the State Options section, change the Sampling options to Single Clock.
In the clock selection area, change the Pod 1 clock to Both Edges.
Close the dialog by clicking OK.
Save the configuration by selecting File then Save As ....
Keysight 16860 Series Portable Logic Analyzer Service Guide